MSG-67 Process simulation: link lost

Cause: No sign of life came from the computer for the configured time. Injected readings remain valid until replaced – so a broken link would be indistinguishable from “nothing has changed”, and the device would keep heating on a frozen temperature. It has therefore ended the simulation, switched all channels off and dropped the main contactor.
Measure: Check the link to the computer. Then switch the simulation on again; the main contactor has to be released explicitly afterwards.